MSO Series
Portable Mixed Signal Digital Storage Oscilloscope
User Manual
MSO7062TD
MSO7102TD
MSO8102T
MSO8202T
WWW.OWON.COM.HK
Zobrazit stránku 0
1 2 3 4 5 6 ... 111 112

Shrnutí obsahu

Strany 1 - User Manual

MSO Series Portable Mixed Signal Digital Storage Oscilloscope User Manual MSO7062TD MSO7102TD MSO8102T MSO8202T WWW.OWON.CO

Strany 2 - Branch:

5 4. Junior User Guidebook This chapter deals with the following topics mainly: Digital Storage Oscilloscope  Introduction to the front pane

Strany 3 - General Warranty

95 Fig.6-2 Wave Form of Gain Measurement Example 3: Capture the Single Signal The digital storage oscilloscope takes the lead in providing

Strany 4 - Table of Contents

96 the signal in conformity with the trigger conditions. If a signal reaches to the set trigger level, one sampling will be made and then displaye

Strany 5

97 Fig.6-4 Signal with Noises Reduce the effect of noises from your signal When focusing on signal itself, the important thing is to reduc

Strany 6

98 Example 5: Examine the Phase shift between two related signals X-Y mode is a very useful when examining the Phase shift of two related signa

Strany 7 - Safety Symbols

99 Example 6: Video Signal Trigger Observe the video circuit of a television, apply the video trigger and obtain the stable video output signal

Strany 8 - Warning:

100 7. F.A.Q 1. In the case of that the oscilloscope is still in the black-screen state without any display after the power is switch on, impleme

Strany 9 - 3. General Characteristics

101 8. Technical Specifications Unless otherwise specified, the technical specifications applied are applicable to the probe with the attenuatio

Strany 10 - 4. Junior User Guidebook

102 Horizontal System Sampling rate range  1S/s~1GS/s half channel*, 1S/s~500MS/s each channel ---- MSO7062TD, 7102TD  1S/s~2GS/s half channel

Strany 11 - Front panel

103 Rise time  ≤5.8 ns (at input, Typical) ----MSO7062TD  ≤3.5 ns (at input, Typical) ----MSO7102TD, 8102T  ≤1.75 ns (at input, Typical) ---

Strany 12 - Control(key and knob) area

104 displacement Trigger Holdoff range 100ns~10s 50% level setting(typical) Input signal frequency ≥50Hz Edge trigger slope Rising、Falling Sensiti

Strany 13

6 4.1 Introduction to the Front Panel and the User's Interface When you get a new-type oscilloscope, you should get acquainted with its front

Strany 14 - User interface introduction

105 Data System Binary system, Decimal system, Hex Digital Filter 0/1/2 optional Setting storage Support USB storage Support General Technica

Strany 15

106 9. Appendix Appendix A: Enclosure Standard Accessories:  Passive probe: 2, 1.2 m, 1:1 (10:1)  OL-16 LA measurement module  CD: x

Strany 16

107 Appendix C: Battery Using Guide Charging the oscilloscope The lithium battery maybe not be charged when delivery. Please charge the batte

Strany 17

7 Control(key and knob) area Fig. 4-2 Keys Overview 1、Menu option setting: F1~F5 2、Switch Switch includes two keys and one knob. Press "

Strany 18

8 idle. "CH1 Vertical", "CH2 Vertical" to adjust the M1, M2 position in Cursor menu when cursor display is on "CH1 Volts/

Strany 19

9 4.2 Digital Storage Oscilloscope User interface introduction Fig. 4-3 Illustrative Drawing of Display Interfaces 1. The Trigger State indicate

Strany 20

10 5. This reading shows the time deviation between the horizontal trigger position and the window center line, which is regarded as 0 in the win

Strany 21

11 waveform of the CH1 channel. If the pointer is not displayed, it shows that the channel is not opened. 22. The positions of two purple dotte

Strany 22

12 10X, 2. Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel. Align the slot in the probe with th

Strany 23 - 4.3 Logic Analyzer

13 Fig. 4-5 Displayed Wave Forms of the Probe Compensation 3. Repeat the steps mentioned if necessary. Fig. 4-6 Adjust Probe How to Set

Strany 24 - How to acquire data

14 Note: When the attenuation switch is set to 1X, the probe will limit the bandwidth of the oscilloscope in 5MHz. If it is needed to use the wh

Strany 25 - Display systems

Nov. 2014 edition Ver1.3.7 Copy Right in this Manual © Lilliput Company. All rights Reserved. The Lilliput's products

Strany 26 - Trigger system

15 Fig. 4-9 Vertical Control Zone 1. Use the button "VERTICAL POSITION" knob to show the signal in the center of the waveform win

Strany 27 - Threshold voltage system

16 Introduction to the Horizontal System Shown as Fig.4-10, there are a button and two knobs in the "HORIZONTAL CONTROLS". The following

Strany 28 - Sampling system

17 Fig.4-11 Trigger Control Zone 1. Press the "TRIG MENU" button and call out the trigger menu. With the operations of the 5 menu s

Strany 29 - Fig. 4-19

18 4.3 Logic Analyzer LA input connection Insert the plug of OL-16 LA module 50P into the LA signal input on front panel and fix two screw. Then 1

Strany 30 - 5. Advanced User Guidebook

19 detected and wait for sample finished. "STOP" for sampling finished。 12、Value indicate current time base。 13、Info windows: different

Strany 31 - Input signal is interrupted

20 Display systems We need only three channels as what we measure is 3 signals. And other channel and bus can be off. In this way the display reso

Strany 32

21 Fig.4-14 Fig. 4-15 Trigger system LA is same as DSO and need to make trigger to synchronize data. The trigger system mainly to set trigger so

Strany 33

22 Fig. 4-16 Threshold voltage system Threshold voltage system is to set high/low of the trigger voltage. The system already fixed the setting fo

Strany 34

23 Sampling system The waveform accuracy reverts from sample data depend on sample rate for measured signals. The waveform reverted in LA is refer

Strany 36

User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope General Warranty Lilliput warrants that the product will be free from defects

Strany 37

25 5. Advanced User Guidebook Up till now, you have already been familiar with the initial operations of the functions of the function areas, b

Strany 38

26  How to set Utility It is recommended that you read this chapter carefully to get acquainted the various measurement functions and other oper

Strany 39

27 from input signal. See Fig.5-2.  Press F1 again to set "DC" mode, both AC and DC components get passed. See Fig.5-3. The wave forms

Strany 40 - Knobs

28 Taking the Channel 1 for example, the operation steps are shown as below: (1). Press the CH1 MENU button and call out the CH1 SETUP menu.

Strany 41 - Set Window

29 Fig. 5-5 Regulation of the Attenuation Ratio of the Probe A List of the Attenuation Coefficient of Probes and the Corresponding Menu Setting

Strany 42 - Zone Window

30 Implementation of Mathematical Manipulation Function The Mathematical Manipulation function is used to show the results of the additive,m

Strany 43 - How to set trigger system

31 Function Menu Setting Instruction FFT ON OFF Turn on FFT function. Turn off FFT function. Source CH1 CH2 DO FFT on Channel l DO FFT on Chan

Strany 44 - Single trigger

32 to select the best window. Type Description Window Rectangle This is the best type of window for resolving frequencies that are very close to t

Strany 45 - Video Trigger

33 Fig.5-9. Blackman window Fig.5-10 Hamming window Fig.5-11. Rectangle window

Strany 46 - Pulse Width Trigger

34 Fig.5-12 Hanning window Quick Tips  If desired, use the zoom feature to magnify the FFT waveform.  Use the default dBV RMS scale to see a

Strany 47 - Alternate trigger

i Table of Contents 1. General Safety Requirements ... 1 2. S

Strany 48 - Menu list:

35 Application of VERTICAL POSITION and VOLTS/DIV Knobs 1. The .VERTIVAL POSITION knob is used to adjust the vertical positions of the wave for

Strany 49 - Pulse) Menu

36 Fig. 5-14 Time Mode Main Time Base Press the F1 menu selection button and choose the Main Time Base. In this case, the HORIZONTAL POSITION

Strany 50 - Slope) menu list:

37 Fig. 5-16 Window Setting Fig..5-17 Set Window disable under FFT mode Zone Window Press the F3 menu selection button and choose Zone Wi

Strany 51

38 Fig. 5-18 Zone Window How to set trigger system When the oscilloscope begins to collect the data and display the wave form depends on a trig

Strany 52

39 The Single Trigger and Alternate Trigger menus are described respectively as follows: Single trigger Single trigger has four modes: edge trigge

Strany 53

40 Coupling AC DC HF LF Block the direct current component. Unblock all components. Block the high-frequency signal and only unblock the low-

Strany 54 - waveform

41 Pulse Width Trigger Pulse trigger occurs according to the width of pulse. The abnormal signals can be detected through setting up the pulse w

Strany 55 - XY Format

42 Slope Trigger Slope trigger sets the oscilloscope as the positive/negative slope trigger within the specified time. The Slope Trigger Menu is

Strany 56

43 mode is used to observe two unrelated signals. You can choose different trigger modes for different channels. The options are as follows: ed

Strany 57 - Recall a stored waveform

44 Fig.5-24 Alternate trigger(Trigger Type: video) Menu Alternate trigger(Trigger Type: video) Menu list: MENU SETTING INSTRUCTION Source CH1

Strany 58 - Fig.5-39 FFT wave saving

ii How to Implement the Automatic Measurement ... 55 How to Implemen

Strany 59 - SYS STAT (System State)

45 Alternate trigger(Trigger Type: Pulse) menu list MENU SETTING INSTRUCTION Source CH1 CH2 Channel 1 as the trigger source. Channel 2 as the tr

Strany 60

46 Condition Rising Edge and less than. Rising Edge and equal to . Falling Edge and more than Negative pulse and less than Negative pulse and

Strany 61

47  Normal: The Normal mode allows the oscilloscope to acquire a waveform only when it is triggered. If no trigger occurs, the oscilloscope keep

Strany 62

48 waveform. You can see the "burrs" on the falling edge of signal is detected with noise when it is turned on. Fig.5-28 Peak Detect mo

Strany 63

49 How to Set the Display System Push down the DISPLAY button and the menu displayed in the screen is shown as Fig.5-31. Fig.5-31 Display Set

Strany 64

50 Fig.5-33 Display in Dots form Persist When the Persist function is used, the persistence display effect of the picture tube oscilloscope can

Strany 65 - MATH FFT

51  The Vertical VOLTS/DIV and the Vertical POSITION knobs of Channel 1are used to set the horizontal scale and position.  The Vertical VOL

Strany 66

52 How to Save and Recall a Wave Form Press the SAVE/RCL button, you can save and call out the waveforms in the instrument. The menu displayed

Strany 67 - How to use Autoscale

53 Fig.5-38 Wave Saving Fig.5-39 FFT wave saving

Strany 68 - Fig.5-51

54 How to Implement the Auxiliary System Function Setting Press the UNTILITY button and the menu is displayed in the screen as Fig.5-40. Fig.5-

Strany 69 - U-DISK COPY

1 1. General Safety Requirements Before any operations, please read the following safety precautions to avoid any possible bodily injury and pre

Strany 70

55 Fig.5-41 SYS STAT Menu After entering into the SYS STAT menu, choose the corresponding function, with the corresponding parameters shown in

Strany 71

56 Overshoot, Preshoot, RiseTime, Fall Time, +Width, -Width, +Duty, -Duty, DelayA B and DelayA B . To use Measure, simply Press Measure button

Strany 72 - Figure: Format setting

57 Average The arithmetic mean over the entire waveform. Vrms The true Root Mean Square voltage over the entire waveform. Rise Time Time that the

Strany 73 - Figure: Format successfully

58 Fig.5-44) Fig.5-44 Automatic Measurement How to Implement the Cursor Measurement Press the CURSOR button to display the cursor measurement

Strany 74 - 5.2 Logic analyzer

59 Delta Absolution value of the difference between two cursors. Cursor 1 Reading of Cursor 1. Time: read time from start position-the horizon

Strany 75 - Low memory

60 Fig.5-47 Wave Form of Cursor Measurement The Cursor Measurement for FFT model: Press the CURSOR button to display the cursor measurement fun

Strany 76

61 When carrying out the cursor measurement, the position of Cursor 1 can be adjusted with the CURSOR1 (VERTICAL POSITION) knob of Channel 1, and

Strany 77 - Fig.5-56

62 Fig.5-50 wave of Freq cursor measurement How to use Autoscale The function is applied to follow-up signals automatically even if the signals c

Strany 78 - Fig.5-57)

63 Fig.5-51: Multi cycles mode single cycle mode Note: Things you should be ware when turning on Autoscale. 1. The symbol is

Strany 79 - Fig.5-58)

64 Horizontal Sale Adjust to the proper division Trigger Type Current Trigger Source Show the minimum number of channels. Trigger Coupling Cur

Strany 80 - Fig.5-60)

2 2. Safety Terms and Symbols Safety Terms Terms in this manual. The following terms may appear in this manual: Warning: Warning indicates

Strany 81

65 Figure: Disk Management of computer 4. Right click 1 or 2 red mark area, choose Format. And system will pop up a warning message, click Yes.

Strany 82 - Fig.5-62

66 6. Formatting process. Figure: Formatting the USB disk 7. Check whether the USB disk is FAT32 with allocation unit size 4096 after formattin

Strany 83

67 Figure: Reload Disk 4. Right click 1 or 2 red mark area, choose Format. Figure: Choose format 5. Set File System FAT32, Cluster size 4096.

Strany 84 - How to set threshold

68 Figure: Apply setting 7. Formatting process Figure: Format process 8. Format the USB disk successfully Figure: Format successfully

Strany 85 - Fig.5-66

69 5.2 Logic analyzer How to set sampling system Sampling system is to set sample rate, storage depth and filter. Different sampling setting will

Strany 86 - How to set display system

70 Listing of corresponding continuance time to different sample rate and storage depth: Sample rate Storage depth Continuance time Sample rate St

Strany 87 - How to set BUS

71 3. Press "F2" till storage depth display as "Deep Memory". 4. Press "F3" till digital filter display as "Non

Strany 88 - Fig.5-71)

72 Function Settings Instructions Signal choice CH00~CH0F CH00-CH0F can be set as trigger resource Slope Rising Falling Either Trigger on the risi

Strany 89 - How to save and recall

73 Fig.5-56:Bus trigger menu Bus trigger menu as below: Function Setting Instruction Source BUS0~BUS3 BUS0 to BUS3 can be set as trigger source C

Strany 90 - Fig.5-74)

74 Fig.5-57:Bus trigger 3. Pattern trigger: Set channel as signal source and make high/low voltage for channel as trigg

Strany 91 - How to search

3 Warning: The channels should adopt common basis during measuring. To prevent short circuits, the 2 probe ground must not be connected to 2 di

Strany 92

75 CH00~CH0F Don't Care Low High No trigger occurs. Trigger occur when signal on the low level Trigger occurs when signal on the high level.

Strany 93 - Fig.5-76.:Search Bus

76 Fig.5-60:Sequential queue trigger Sequential queue trigger function as below: Function S

Strany 94 - How to review setting info

77 Then sequential queue trigger setting finished (refer to Fig.5-61) and data ready for acquisition. Fig.5-61:Sequential queue trig

Strany 95

78 (DEC) Add Add the code type to the queue Delete Delete the code type from the queue Setting operation are same as sequential trigger setting.

Strany 96

79 Duration 10ns~50s The duration width can be set from 10ns (1-2-5) to 10.00us (1-2-5) according to the sampling rate from high to low. Qualif

Strany 97 - How to set Utility

80 measurement. For example, if measure signal is LVCMOS1.8V and set threshold as "CMOS/(2.5V)" then all the measurement data will becom

Strany 98 - 6. Demonstration

81 Threshold voltage setting finished (refer to Fig.5-67). Fig.5-67:Threshold custom setting How to set display system Dis

Strany 99 - Metering Circuit

82 display ON OFF BUS or Channel on Bus or Channel off contrast Increase Increase panel display contrast contrast Decrease Decrease panel disp

Strany 100

83 Fig.5-70:Bus menu BUS setting menu as below: Function Setting Instruction BUS BUS0~BUS3 Choose BUS for operating Channel CH0F~CH00 1X111111 1

Strany 101

84 Fig.5-71:Bus setting How to measure Measurement can take auto measure for values of 4 BUS synchronously. Press "me

Strany 102

4 3. General Characteristics Digital Storage Oscilloscope  Model Bandwidth Sample Rate MSO7062TD 60MH

Strany 103

85 display setting, threshold setting and trigger setting. And it is possible to save 10 groups of settings. Waveform storage and setting storage

Strany 104

86 Fig.5-74:Waveform saving How to use USB flash disk to storage Use USB flash disk is to storage acquired data. Insert U disk

Strany 105 - 7. F.A.Q

87 Fig.5-75:Search trigger position 2. Search specified value in BUS Searching BUS menu as below: Function Setting Instruction BUS BUS0~BUS3 Choo

Strany 106 - Digital Storage Oscilloscope

88 Fig.5-76.:Search Bus 3. Searching target is pattern: pattern refers to the complex of different channel according to hig

Strany 107

89 Fig.5-77.:Search Code type How to review setting info You can choose system information display to be "On" or &quo

Strany 108

90 Fig.5-79:Cursor measurement menu Cursor measurement time menu as below: Function Setting Instruction Increment Time Frequ

Strany 109 - Logic analyzer

91 Fig.5-80:Cursor time measurement Cursor measure position menu as below: Function Display Instruction M1-M2 Position The

Strany 110

92 Fig.5-81:Cursor position measurement How to set Utility Utility function includes recall factory, Language, Carry. Utilit

Strany 111 - 9. Appendix

93 6. Demonstration Notice: The following is to take MSO7102T model as an example. Example 1: Measurement of Simple Signals The purpose of thi

Strany 112

94 Example 2: Working out the Gain of the Amplifier in the Metering Circuit The purpose of this example is to work out the Gain of an Amplifier in